Professional guidelines for cleaning, sterilizing and maintaining your surgical instruments.
Guidelines for cleaning, sterilizing and maintaining surgical instruments. Use of distilled or deionized water and a neutral pH cleaning solution is recommended for all procedures.
Immediately after surgery, rinse instruments under warm (not hot) water. It may be helpful to use a nylon toothbrush to rinse the lock boxes and joints of the instrument. Be sure to remove all blood, body fluids, and tissue.
If you do not clean your instruments immediately after rinsing, submerge them in a solution of water and neutral pH (7) detergent. Never place instruments in saline solution, as it may cause corrosion and irreversible damage.
We recommend you manually clean micro and delicate instruments. Process instruments in a cleaner for the full recommended cycle time — usually 5–10 minutes. Place instruments in the open position. Ensure sharp blades (scissors, knives, osteotomes, etc.) do not touch other instruments. All instruments must be fully submerged. Do not place dissimilar metals (stainless, copper, chrome plated, etc.) in the same cleaning cycle. Change solution frequently, at least as often as recommended by the manufacturer. Rinse with distilled or deionized water after ultrasonic cleaning.
Follow the manufacturer's recommendations. Make sure instruments are lubricated after the last rinse cycle and before the sterilization cycle.
Most manufacturers recommend ultrasonic cleaning as the best method for instruments with hinges, locks, and moving parts. If ultrasonic cleaning is not available, observe the following steps:
If instruments are to be stored, let them air dry and store them in a clean, dry environment.
If instruments are to be reused or autoclaved:
Never lock an instrument during autoclaving — steam cannot reach metal-to-metal surfaces and heat expansion may cause cracks. Do not overload the autoclave chamber. Place a towel on the bottom of the pan to absorb excess moisture and reduce wet packs. Ensure sterilization towels have no detergent residue and are neutral pH (7).
Most cold sterilization solutions render instruments sterile only after a 10-hour immersion, which can be more detrimental than a 20-minute autoclave cycle. If disinfection only is needed, cold sterilization is recommended as disinfection takes place in about 10 minutes.
Remember: Sterile means no living organism survives; Disinfected means basically clean. Always use the proper sterilization and cleaning technique.
Need advice on instrument care? Contact our team.
Contact UsProfessional guidelines for cleaning, sterilizing and maintaining your surgical instruments.
Guidelines for cleaning, sterilizing and maintaining surgical instruments. Use of distilled or deionized water and a neutral pH cleaning solution is recommended for all procedures.
Immediately after surgery, rinse instruments under warm (not hot) water. It may be helpful to use a nylon toothbrush to rinse the lock boxes and joints of the instrument. Be sure to remove all blood, body fluids, and tissue.
If you do not clean your instruments immediately after rinsing, submerge them in a solution of water and neutral pH (7) detergent. Never place instruments in saline solution, as it may cause corrosion and irreversible damage.
We recommend you manually clean micro and delicate instruments. Process instruments in a cleaner for the full recommended cycle time — usually 5–10 minutes. Place instruments in the open position. Ensure sharp blades (scissors, knives, osteotomes, etc.) do not touch other instruments. All instruments must be fully submerged. Do not place dissimilar metals (stainless, copper, chrome plated, etc.) in the same cleaning cycle. Change solution frequently, at least as often as recommended by the manufacturer. Rinse with distilled or deionized water after ultrasonic cleaning.
Follow the manufacturer's recommendations. Make sure instruments are lubricated after the last rinse cycle and before the sterilization cycle.
Most manufacturers recommend ultrasonic cleaning as the best method for instruments with hinges, locks, and moving parts. If ultrasonic cleaning is not available, observe the following steps:
If instruments are to be stored, let them air dry and store them in a clean, dry environment.
If instruments are to be reused or autoclaved:
Never lock an instrument during autoclaving — steam cannot reach metal-to-metal surfaces and heat expansion may cause cracks. Do not overload the autoclave chamber. Place a towel on the bottom of the pan to absorb excess moisture and reduce wet packs. Ensure sterilization towels have no detergent residue and are neutral pH (7).
Most cold sterilization solutions render instruments sterile only after a 10-hour immersion, which can be more detrimental than a 20-minute autoclave cycle. If disinfection only is needed, cold sterilization is recommended as disinfection takes place in about 10 minutes.
Remember: Sterile means no living organism survives; Disinfected means basically clean. Always use the proper sterilization and cleaning technique.
Need advice on instrument care? Contact our team.